NeuroTrac® MyoPlus4 Pro
- New and Improved touch-screen with full colour display. The new user friendly interface makes it a self-explanatory unit with the interactive learning on the go, including electrode placement, running EMG graph as well as many other helpful tools.
- Pre-set programmes for Incontinence treatment, Sports and muscular development, neuro-muscular Rehabilitation.
- Designed for Physiotherapist in mind to be used as a portable tool. Suitable as a personal trainer for in-clinic or home use (under guidance).
- 2 channel EMG (Biofeedback) combined with 4 channels of NMES (stimulation).
- 2 channel ETS (EMG triggered stimulation) with stimulation on 4 channels.
- Simple single-phase or multiphase operation: multiphase allows EMG/ETS/STIM to be combined in one programme.
- Multilingual LCD screen and voice prompts for Biofeedback.
- Manual and Automatic threshold control.
- EMG Biofeedback between stimulation (muscle activity and tiredness indication when using muscle stimulation).
- Stand alone device or used in conjunction with established multilingual software.
- Now with Bluetooth connectivity! The unit sends up to 10 metres of wireless data to PC software.
- For home users: Produces and prints (with the optional PC Software) the comprehensive patient progress reports on a day to day basis providing average readings (2 channels) on work/rest, peak, onset and muscle release times, work and rest deviations and the average mA current.
- Used as a learning tool for both patient and clinician with emphasis on improving treatment techniques in the use of EMG and Neuromuscular stimulation.
- 4 x AA batteries {No mains required}.
Pelvic Floor: P01-P19 | Sports: P01-P21 |
Rehab: P01-P08 |
CUSTOM |
P1 - Pelvic floor pain |
P1 - Warm Up |
P1 - Muscle growth |
EMG Custom x 2 STIM Custom x 2 ETS Custom x 2 P&A Custom x 5 P+A Custom is suitable for a tailor made programme to be set up with Passive STIM and Active EMG / ETS phases, all in one programme. |
Protocols | Author | Download |
---|---|---|
Treatment of the Hemiplegic Shoulder (Stroke) through Biofeedback: Case Study | Verónica Bargueño, Juan Nicolás Cuenca & Eric Lazar | |
F.E.S. in Stroke | Juan Nicolás Cuenca & Eric Lazar | |
Urinary Incontinence | Julia Herbert Grad. Dip. Phys. MCSP SRP Specialist Continence Physiotherapist | |
Stress Urinary Incontinence | Anna Pawlaczyk Specialist in Gynecology | |
Stress Urinary Incontinence, using EMG Triggered Stimulation - NEW | Maciej Kisiel, Henryk Konon MD | |
Pelvic Floor Muscle Activity in Sitting & Standing Postures | R Saphod, C Maher and C Richardson | |
Fecal Incontinence | Pirkko Raivio Specialist Continence Physiotherapist | |
Incontinence Case Reports | Pirkko Ravio Specialist Continence Physiotherapist | |
General Pain Relief with TENS | Dr. Deirdre M. Walsh Phil, B. Physio, PgCUT, MISCP, MCSP Co-ordinator, Rehabilitation Sciences Research Group | |
Patellofemoral Pain | James Selfe. PhD. MA. GD Phys. MCSP. SRP. Lecturer Chartered Physiotherapist. | |
Prolapse: Pelvic Floor Muscle Training for Women | Karen Radford, MSc, MCSP and Funmi Odofin, BSc, PG Cert. MCSP |
EMG (Biofeedback)
- Dual channel EMG
- EMG Range: 0.2 to 2000 μV RMS (continuous)
- Sensitivity: 0.l μV RMS
- Accuracy: 4% of μV reading +/-0.3 μV at 200 Hz
- Selectable Bandpass filter - 3db Bandwidth,
- a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% - Reading below 235 microvolts 10 Hz +/-3 Hz to 370 Hz +/- 10% - Reading above 235 microvolts
- b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
- Notch filter: 50 Hz (Canada 60Hz) - 33 dbs (0. 1% accuracy)
- Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
- Batteries: 4 x 1.5V ( AA battery)
- Optional Charger: CE approved.
- Work / Rest periods: 2-99 seconds
- Number of Trials: 2-99
STIM (Neuromuscular Stimulation)
- Four channel Stimulator
- Amplitude: 0-90 mA into 500 Ohm load - actual mA will tend to be less
than indicated due to Electrode impedance: at 1000 Ohms load (Electrodes
in poor condition) the maximum will be limited to 75 mA, at 1500 Ohms
load the maximum will be limited to 50 mA. - Type: Constant current, maximum output
voltage 70 Volts +5 / -10 Volts - Waveform: Symmetrical, rectangular,
bi-phasic with net zero DC current - Pulse width selection: 50-450 μS (2% accuracy)
- Pulse rate selection: 2-100 Hz (2% accuracy)
- Work / Rest periods: 2-99 seconds
- Time 1 - 99 minutes
- Ramp up time: 0.1 - 9.9 seconds
- Preset and user programmable treatment Programs
- Automatic output shut off with detection of open electrode above 0.5 mA
Battery: Low battery indication at 5.4 volts +/- 0.2 volts, automatic shut off when voltage dropps below the low indication. Replace the batteries immediately! The device goes off automatically when not in use (energy saving): for example when in settings and no key pressed over 1 minute, when in stimulation mode (home screen) and all channels are 0mA.
Expected average battery set life [of standard 800 mAh, alkaline]:
a. MyoPlus2: 12-16 hours in STIM, 25 hours in EMG mode.
b. MyoPlus4: 8-16 hours in STIM, 25 hours in EMG mode.
Environmental Conditions for use: +10 to +30 degrees Centigrade. 0-90% Humidity.
Environmental conditions for storage & transport: -10 to +50 degrees Centigrade. 0-90% Humidity.
Physical dimensions: Length 150 mm, Width 89 mm, Depth 35 mm.
Weight: MyoPlus2 device: 0.16 Kg (without batteries).